pH Sensor PD1R1
Hach pHD Online Process pH Sensor - pH Sensor for Clean WaterThe smart choice for accurate and reliable online process pH measurementClean Water Online Process pH Sensor - pHD Technology, Glass pH Electrode, PPS Housing, Convertible Mount, 4.5m CableBenefitsExceptional Process pH Sensor Performance
Introduction
Hach pHD Online Process pH Sensor - pH Sensor for Clean Water
The smart choice for accurate and reliable online process pH measurement
Clean Water Online Process pH Sensor - pHD Technology, Glass pH Electrode, PPS Housing, Convertible Mount, 4.5m Cable
Benefits
Exceptional Process pH Sensor Performance with the Differential Electrode pHD Measurement Technique
This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These process pH sensors provide greater reliability, resulting in less downtime and maintenance.
Extended Working Life with the Replaceable Salt Bridge/Protector
The unique, replaceable salt bridge holds an extraordinary volume of buffer to extend the working life of the sensor by protecting the reference electrode from harsh process conditions. The salt bridge simply threads onto the end of the sensor if replacement is needed.
Patented Technology
The former GLI, now a Hach Company brand, invented the Differential Electrode Technique for pH measurement in 1970. The pHD™ sensor series (U.S. Patent Number 6395158B1, dated May 28, 2002) takes this field-proven technology to a new level.
Lower Maintenance Needs with the Double Junction Salt Bridge
The double junction salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.
Reliability with Built-in Encapsulated Preamp
Encapsulated construction protects the sensor’s built-in preamp from moisture and humidity, ensuring reliable sensor operation. The preamp in the pHD analog sensor produces a strong signal, enabling the sensor to be located up to 1000 m (3280 ft.) from the analyzer.
Versatile Mounting Styles
Sensors are available in four mounting styles—convertible, insertion, immersion, and sanitary.
Specifications
Technical Attributes
Accuracy | ± 0.02 pH |
Body material | PPS |
Cable Connection | Analog |
Cable Length | 4.5 m 5-conductor (plus two isolated shields) cable with XLPE (cross-linked polyethylene) jacket; rated to 150 °C) |
Cable Length | 4.5 m |
Calibration Method | Two point automatic, one point automatic, two point manual, one point manual. |
Communication | MODBUS |
Diameter | 34.9 mm |
Drift | 0.03 pH per 24 hours, non-cumulative |
Electrode Type | General Purpose |
Flow | max. 3 m/s |
Length | 196.3 mm |
Material | Ground Electrode: Titanium |
Mounting | Convertible |
Operating Temperature Range | -5 - 95 °C |
Pressure Range | max. 10.7 bar Sensor only (pressure range of mounting hardware to be checked separately) |
Range | -2.0 - 14.0 pH |
Repeatability | ± 0.05 pH |
Sensor Thread | 1" NPT at both ends |
Special Notes | pH-Wide Range |
Temperature Accuracy | ± 0.5 °C |
Temperature Compensation | automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature, additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 ° |
Temperature Compensation | automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature, additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 ° |
Temperature Sensor | NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout |
Warranty | 12 months |
Weight | 0.9534 kg |
What's included? | Includes: sensor with 15 ft cable and manual |