Home >> Products >> HACH >> Portable

Conductivity Probes CDC401

Product Model: IntelliCAL
Product Number: CDC40103

HACHIntelliCALⓇ CDC401 is a digital, graphite, 4-pole conductivity cell with a temperature sensor. The laboratory version is available with a 3 or 1 m cable. The CDC401 is ideal for measuring electrical conductivity, salinity, resistivity, or total dissolved solids (TDS) in wastewater,

Introduction

HACH IntelliCALⓇ CDC401 is a digital, graphite, 4-pole conductivity cell with a temperature sensor. The laboratory version is available with a 3 or 1 m cable. The CDC401 is ideal for measuring electrical conductivity, salinity, resistivity, or total dissolved solids (TDS) in wastewater, drinking water, or general water quality applications.

Benefits

Hach IntelliCAL digital probes provide ultimate traceability in conductivity measurement history

Stored time and date stamp for each measurement, operator and sample ID, calibration history, parameter, and probe serial number.

IntelliCAL CDC401 digital conductivity probes alert the user when re-calibration is needed

No guesswork needed to provide the most reliable and accurate results with each measurement.

IntelliCAL CDC401 digital conductivity probes can be moved between meters without the need to re-calibrate or re-enter measurement settings

Ideal for a multi-user environment with multiple HQD series Laboratory and Portable meters.

Specifications

Technical Attributes

AccuracyCond: ±0.5% of reading|TDS: ±0.5% ±1 digit|Salinity: ±0.1, ±1 digit
Cable Length1 m (3.28 ft)
Electrode TypeConductivity Cell;
4 Poles - Graphite
InstrumentSensor & Electrodes
Kit?No
Length184 mm (7.24 in.)
Method TypeLaboratory: 4 Poles - Graphite
ParameterConductivity
Probe TypeStandard
Product KitModel: CDC401
Accessories Included: None
RangeConductivity: 0.0 µS/cm - 200 mS/cm
TDS: 0.00 mg/L - 50.0 g/L as NaCl
Salinity: 0 - 42 ppt or ‰
Resistivity: 2.5 Ωcm - 49 MΩcm
Resolution0.01/0.1 (5 digits max.)
Sample depth45 mm (1.77 in.)
Sensor materialNoryl
Sensor Type4-poles conductivity probe, Graphite, k = 0.40 cm-1
Temperature Accuracy±0.3 °C (±0.54 °F)
Temperature Range-10 - 110 °C (14 - 230 °F)
Temperature Resolution0.1 °C (0.18 °F)
Test requirementsParameter Needed: Conductivity
Minimum Sample Depth (mm): 45
ThermistorATC
Warranty12 months
Weight0.1 kg
What's included?IntelliCAL CDC401 Laboratory Conductivity Cell, 1 m cable, Test certificate, and Basic User Manual.