Home >> Products >> GLI

PRO Series Transmitter, pH/ORP

Product Model: PRO-P3
Product Number: PRO-P3A1N

The economical PRO-P3 Transmitter is easy to use with its menu-guided user interface. The NEMA 4X enclosure can be panel, wall, pipe, or integral sensor mounted. The interface is identical to GLI's 33-series analyzers. Versatile 2, 3, or 4-wire hookup capability. Isolated 4 - 20 mA output.

Introduction

The economical PRO-P3 Transmitter is easy to use with its menu-guided user interface. The NEMA 4X enclosure can be panel, wall, pipe, or integral sensor mounted. The interface is identical to GLI's 33-series analyzers. Versatile 2, 3, or 4-wire hookup capability. Isolated 4 - 20 mA output. The transmitter is also passcode protected.

Benefits

Multiple Measurements (pH, ORP, temperature, analog output value)

Multiple Language Capability

Versatile Sensor Compatibility

Auto/Manual Temperature Compensation

Simple Interactive Diagnostics

Technical Attributes

Accuracy± 0.1 % of span, at 25 °C
Calibration Method2-point Buffer (for pH only)
1-point Sample (for pH or ORP)
Communication0/4 to 20 mA with 0.004 mA (12-bit) resolution (during calibration, the analog output is automatically held at the last measured value and, upon completion, returned to its active state)
Dimensions (H x W x D)95 mm x 95 mm x 60 mm
Display2-line by 16 character LCD
LoadMaximum permissible load dependent on power supply voltage, transmitter hookup arrangement, and wire resistance
MaterialEnclosure: Polycarbonate
Measuring Range ORP-2100 - 2100 mV
Measuring Range pHSelectable Ranges:
-2.0 - 14.0 pH
-2.00 - 14.00 pH
Measuring Range Temperature-20 to 200 °C
Memory BackupAll user settings are retained indefinitely without battery backup
MountingPanel, wall, pipe, or integral sensor mounting
Operating Temperature Range-20 - 60 °C at 0 - 95% RH (non-condensing)
ParameterpH/ORP
Power Requirements (Voltage)12 - 30 V DC
Relative Humidity0 - 95 % non-condensing
Repeatability± 0.05 % of span, at 25 °C
Response Time T901 - 60 s with output filter setting of zero
Sensitivity± 0.05 % of span, at 25 °C
Special NotesThe measured pH (or ORP) and temperature can be displayed separately or together on one screen. Both analog output values are
shown together on one screen
Storage Conditions-30 to 70 °C 0-95% Relative humidity, non-condensing
Temperature Compensation-10 - 110 °C
Automatic with selection for NTC 300 Ω, Pt1000 or Pt100 or manually fixed at a user-entered temperature; additional selectable temperature correction factors (ammonia, morpholine or user-set pH/°C linear slope)
Temperature Drift0.02 % of span per °C (zero and span), at 25 °C
Warranty12 months
Weight0.280 kg
What's included?Controller, User Manual